Quick answer. SpCO is a noninvasive estimate of carboxyhemoglobin. SpMet is a noninvasive estimate of methemoglobin. They are different analytes, and they answer different clinical questions. On Masimo rainbow 8-lambda SpCO sensors, both parameters can be available on the same sensor when the monitor has them enabled. The reading is still limited by the device, and a laboratory CO-oximeter is the reference method used in the IFU accuracy tables.12
What each parameter is
Masimo LAB-9373E, the directions for use for RD rainbow 8-lambda SpCO adhesive sensors, indicates those sensors for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO) and methemoglobin saturation (SpMet).1 The LNCS-II rainbow DCI and DCIP 8-lambda SpCO IFU, LAB-10572A, uses the same parameter set and states that the sensor is limited by the parameters on the device.2
In plain terms:
- SpCO is the pulse CO-oximeter's estimate of carboxyhemoglobin, the form of hemoglobin bound to carbon monoxide.
- SpMet is the pulse CO-oximeter's estimate of methemoglobin, the form of hemoglobin in which the iron is in the ferric state and does not carry oxygen normally.
A conventional two-wavelength pulse oximeter reports SpO2. It does not separately report SpCO or SpMet. Rainbow 8-lambda SpCO sensors are built to add those dyshemoglobin readings when the monitor is licensed for them.1
Why the distinction matters in EMS
Carbon monoxide exposure and methemoglobinemia can both leave a patient hypoxemic, and both can make a standard SpO2 reading harder to interpret, but they are not the same problem and they are not treated the same way.
The U.S. Food and Drug Administration's 510(k) summary for Rainbow Adhesive Pulse CO-Oximeter Sensors (K071024) states that those sensors are indicated for continuous noninvasive monitoring of SpO2, pulse rate, carboxyhemoglobin (SpCO), and/or methemoglobin (SpMet) saturation.3 FDA Unique Device Identification records for the later RD rainbow 8-lambda adhesive models 4035 and 4037, and for the adult and infant models 4034 and 4036, list K071024 as the premarket submission.4
A 2006 human volunteer study in Anesthesiology evaluated the Masimo Rad-57 against laboratory CO-oximeters after induced carboxyhemoglobin (inspired carbon monoxide) and induced methemoglobin (intravenous sodium nitrite). The published abstract reports that the Rad-57 measured carboxyhemoglobin with an uncertainty of +/-2% in the range 0-15%, and methemoglobin with an uncertainty of 0.5% in the range 0-12%.5 That paper is a controlled volunteer study on the Rad-57, not a field trial and not a validation of every later sensor family. LAB-9373E still compares SpCO and SpMet accuracy to a laboratory CO-oximeter, and it lists conditions that can make SpCO or SpMet readings inaccurate, including low perfusion, motion, extreme hemoglobin levels, elevated methemoglobin (which can make SpCO inaccurate), and low arterial oxygen saturation or elevated methemoglobin (which can keep SpCO from being provided).1
The clinical takeaway is narrow and sourced: SpCO is the parameter you look at when the question is carbon monoxide exposure; SpMet is the parameter you look at when the question is methemoglobin. A sensor that is capable of both still only reports what the monitor is licensed and able to display.12
Which live sensors read SpCO and SpMet
| Family | Live parts | Parameters (when enabled) | Patient cable | Source |
|---|---|---|---|---|
| LNCS-II rainbow 8-lambda SpCO reusable | 4067, 4068 | SpO2, pulse rate, SpCO, SpMet | RD rainbow SET (for example 4072, 4073, 4792). Not rainbow RC. | PLM-10530C, LAB-10572A62 |
| RD rainbow 8-lambda SpCO adhesive | 4035 (Pdt, 10-50 kg), 4037 (Neo, under 3 kg or over 30 kg) | SpO2, pulse rate, SpCO, SpMet. LAB-9373E does not list an SpCO accuracy figure for the Neo sensor under 3 kg. | RD rainbow SET. Not rainbow RC. | LAB-9373E, PLM-10530C16 |
| rainbow DCI / DCIP reusable | 2696, 2697 | SpO2, SpCO, SpMet (as titled on the live parts) | rainbow RC, not RD rainbow SET | PLM-10530C6 |
| rainbow R-series adhesive (R25, R20, R25-L, R20-L) | 2221, 2222, 2219, 2220 | SpO2, SpCO, SpMet (weight limits differ by part; R25-L lists SpCO on adults over 30 kg) | rainbow RC, not RD rainbow SET | PLM-10530C and the live part pages6 |
| rainbow direct connect | 2201, 2069, 2407 | SpO2, SpCO, SpMet on the live titles | None. The sensor plugs into the monitor. | Stryker LIFEPAK 15 guide, direct-connect section, and the live part pages7 |
Adult and infant RD rainbow 8-lambda adhesive parts are also live: 4034 (adult, over 30 kg) and 4036 (infant, 3-10 kg and 10-30 kg). They use the same RD rainbow SET cable family as 4035 and 4037.
2296 is not a sensor. It is a SpCO parameter upgrade for a Rad-57 that does not already include SpCO. It does not add a SpMet sensor, and it does not change the cable family of the sensor you already have.
Full family lists are on the rainbow sensors collection and the rainbow cables collection.
What this is not
- This is not a comparison with RD rainbow 12-lambda or RD rainbow SET-2 SpHb sensors. Those families add total hemoglobin (SpHb) and are a different part line. They are not in the live rainbow sensors collection reviewed for this article, and they are not a substitute name for 4035 or 4067.
- This is not a treatment protocol, and it does not replace a blood sample or a laboratory CO-oximeter when one is required.
- LAB-9373E does not list SpCO accuracy for the Neo sensor on patients under 3 kg. Do not treat 4037 as a sourced SpCO sensor in that weight range.1
Questions that come up on scene
If the sensor reads SpCO, do I still need a SpMet sensor?
On the 8-lambda SpCO families in the table above, the same sensor is indicated for both parameters. Whether you see SpMet depends on the monitor license and software, not on buying a second 8-lambda SpCO sensor.12
Can a rainbow RC cable and an RD rainbow SET cable be swapped for this?
No. The sensor family decides the cable family. LNCS-II and RD 8-lambda SpCO sensors use RD rainbow SET. rainbow DCI and R-series adhesives use rainbow RC.6 See Which cable fits Masimo rainbow 4067/4068? for the 4067/4068 case.
References
- Masimo. Directions for Use, RD rainbow 8-lambda SpCO Adhesive Pulse CO-Oximetry Sensors. LAB-9373E-eIFU. https://techdocs.masimo.com/globalassets/techdocs/pdf/lab-9373e-eifu.pdf. Opened 2026-11-21. Indications, weight ranges, SpCO and SpMet accuracy table, and listed conditions that can affect readings.
- Masimo. Directions for Use, LNCS-II rainbow DCI and DCIP 8-lambda SpCO. LAB-10572A. https://techdocs.masimo.com/globalassets/techdocs/pdf/lab-10572a-eifu.pdf. Opened 2026-11-21.
- U.S. Food and Drug Administration. 510(k) summary, Rainbow Adhesive Pulse CO-Oximeter Sensors, K071024. https://www.accessdata.fda.gov/cdrh_docs/pdf7/K071024.pdf. Opened 2026-11-21.
- U.S. National Library of Medicine / FDA Unique Device Identification. Model 4034, Primary GTIN 10843997010860; model 4035, 10843997010877; model 4036, 10843997010884; model 4037, 10843997010891. Each record lists premarket submission K071024. Accessed via openFDA and AccessGUDID on 2026-11-21. Example: https://accessgudid.nlm.nih.gov/devices/10843997010860.
- Barker SJ, Curry J, Redford D, Morgan S. Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study. Anesthesiology. 2006;105(5):892-897. PMID 17065881. https://pubmed.ncbi.nlm.nih.gov/17065881/. DOI: 10.1097/00000542-200611000-00008. Abstract values cited above.
- Masimo. SET and rainbow Sensor and Cable Compatibility. PLM-10530C-0323. March 2023. https://www.masimo.co.uk/siteassets/uk/documents/pdf/plm-10530c-sales-tool-sensor-and-cable-compatibility-british.pdf. Opened 2026-11-21.
- Stryker. LIFEPAK 15 accessories guide. February 2025. Direct connect Rainbow SpO2, SpCO, SpMet patient sensors, page 9. https://www.stryker.com/content/dam/stryker/ems/resources/accessory-catalogs/lifepak15_accessory_catalog.pdf. Opened 2026-11-21.