Quick answer. Masimo LNCS-II rainbow 8-lambda SpCO reusable sensors 4067 (adult) and 4068 (pediatric) take an RD rainbow SET patient cable, not a rainbow RC cable. Masimo PLM-10530C lists that family as compatible with RD rainbow SET (RD15) and, for SpO2 only, with a plain RD SET cable. It lists the family as not compatible with rainbow RC.1 The family IFU, LAB-10572A, says the sensors are used with RD/LNC-II rainbow patient cables, and that the parameters you get are limited by the device.2 On a 20-pin monitor, live RD rainbow SET examples in this catalog are MD20-04 EMS (4792), MD20-05 (4072) and MD20-12 (4073). RA25-04 EMS (4798) is an RD rainbow SET 25-pin right-angle cable for a 25-pin monitor connector, not a 20-pin connector.
A rainbow RC cable such as RC-4 EMS (4481) is the wrong family for 4067 and 4068.
What 4067 and 4068 actually are
4067 is the LNCS-II rainbow DCI 8-lambda SpCO adult reusable sensor. 4068 is the LNCS-II rainbow DCIP 8-lambda SpCO pediatric reusable sensor. LAB-10572A covers this LNCS-II rainbow DCI/DCIP 8-lambda SpCO family and states that the sensors can read functional oxygen saturation (SpO2), pulse rate, carboxyhemoglobin (SpCO) and methemoglobin (SpMet), but only when those parameters are enabled on the monitor.2 PLM-10530C's software table lists software 7.1 or higher for SpCO on this family when it is used with an RD rainbow SET cable.1
These are not standard LNCS SpO2 sensors, and they do not use Red LNC or LNC cables.1
The cable rule from Masimo PLM-10530C
| Sensor family | Live parts | RD rainbow SET (RD15) | rainbow RC (M15) | Plain RD SET (RD15) |
|---|---|---|---|---|
| LNCS-II rainbow 8-lambda SpCO reusables | 4067, 4068 | Yes, including SpCO and SpMet when the monitor is licensed | No | Yes, SpO2 only |
Source: Masimo, SET and rainbow Sensor and Cable Compatibility, PLM-10530C (03/2023), row "LNCS-II rainbow 8-lambda SpCO Reusables."1 The US chart PLM-10747D shows the same Y/N pattern for this row.3
The practical test is the sensor-end connector. 4067 and 4068 use the RD15 connector on an RD rainbow SET cable. rainbow RC cables use an M15 connector and are for a different sensor family, including rainbow DCI reusables such as 2696 and rainbow R-series adhesives such as 2221.1
Which live cables to reach for
If the monitor connector is 20-pin, use an RD rainbow SET MD20 cable:
- 4072, RD rainbow SET MD20-05, 5 ft
- 4073, RD rainbow SET MD20-12, 12 ft
- 4792, RD rainbow SET MD20-04 EMS, 4 ft
If the monitor connector is 25-pin, 4798 (RA25-04 EMS) is the live RD rainbow SET 25-pin right-angle EMS cable in this catalog. Do not assume a 20-pin MD20 cable will fit a 25-pin port, or the other way around.
A plain RD SET cable can share the RD15 sensor connector, but PLM-10530C limits that pairing to SpO2. It will not carry SpCO or SpMet.1
See the rainbow cables collection for the live RD rainbow SET and rainbow RC groups, and the rainbow sensors collection for the sensor families those cables serve.
Stryker LIFEPAK 15: use the printed numbers, and do not invent a dual map
Stryker's LIFEPAK 15 accessories guide (02/2025), page 7, prints a heading "Masimo SET RC patient cables" over Stryker numbers 11996-000510 (5 ft), 11996-000511 (12 ft) and 11996-000543 (4 ft EMS). Immediately under that block it prints "Masimo RD SET patient cable compatible Rainbow SpO2, SpCO, SpMet sensors" and lists 11996-000519 (adult reusable) and 11996-000520 (pediatric reusable), plus adhesive numbers 11996-000515 through 11996-000518.4
11996-000519 is also the catalog number on Stryker's shop page for the LNCS-II rainbow DCI 8-lambda SpCO adult reusable sensor. 11996-000520 is the matching pediatric DCIP listing.5 Those names match Masimo 4067 and 4068. No opened Stryker page in this review printed the Masimo numbers 4067 or 4068 next to those Stryker numbers.
This article does not treat the "SET RC" heading as proof that 4067 or 4068 take a rainbow RC cable. Masimo's own compatibility chart pairs that sensor family with RD rainbow SET, not rainbow RC.1 It also does not invent a public dual-number line such as "11996-000510 = 4072." A public OEM page that prints both the Stryker cable number and the Masimo cable number on one line was not found. If you stock LIFEPAK 15 cables by the Stryker number, confirm the cable in hand against the monitor connector and the sensor family before you restock.
The LIFEPAK 15 Masimo spoke already groups 4067 and 4068 with 4072, 4073 and 4792.
Zoll X Series: name-matched sensors, no invented cable map
Zoll's pulse oximetry list and the 8000-000862 product page list 8000-000862 as "LNCS-II Rainbow DCI 8-lambda SpCO Adult Sensor, 1/Box," compatible with X Series and X Series Advanced. The same list names 8000-000863 as "Masimo LNCS-II Rainbow SpCO Sensor, Pediatric, 3 ft."6 Those names match Masimo 4067 and 4068. The opened Zoll pages do not print 4067 or 4068. This article does not add a Zoll-to-Masimo cable dual number, because no Zoll cable page was opened that printed both the Zoll cable number and the Masimo cable number on one line.
The Zoll X Series and Propaq MD spoke and the Masimo part number cross-reference are the live place to look up the OEM numbers that have already been published on those pages.
What this guide does not claim
- It does not claim that a plain RD SET cable will read SpCO or SpMet on 4067 or 4068.1
- It does not claim Propaq MD approval or a prohibition for RD rainbow SET cables. That was not verified on the Zoll pages opened for this guide.
- It does not map 4798 to a Stryker LIFEPAK 35 number. That dual map was not verified against a LIFEPAK 35 OEM document in this review.
Questions that come up on the bench
Can I use the rainbow RC cable already on the monitor?
Not with 4067 or 4068. PLM-10530C marks rainbow RC as not compatible with LNCS-II rainbow 8-lambda SpCO reusables.1 A rainbow RC cable is the right family for rainbow DCI reusables and R-series adhesives, not for this LNCS-II 8-lambda pair.
Why does a connected 4067 show SpO2 but not SpCO?
Two separate limits apply. A plain RD SET cable can connect and still be limited to SpO2.1 Even with an RD rainbow SET cable, SpCO and SpMet have to be enabled on the monitor, and PLM-10530C lists software 7.1 or higher for SpCO on this family.12
References
- Masimo. SET and rainbow Sensor and Cable Compatibility. PLM-10530C-0323. March 2023. https://www.masimo.co.uk/siteassets/uk/documents/pdf/plm-10530c-sales-tool-sensor-and-cable-compatibility-british.pdf. Opened 2026-11-21.
- Masimo. Directions for Use, LNCS-II rainbow DCI and DCIP 8-lambda SpCO. LAB-10572A. https://techdocs.masimo.com/globalassets/techdocs/pdf/lab-10572a-eifu.pdf. Opened 2026-11-21.
- Masimo. SET and rainbow Sensor and Cable Compatibility (US). PLM-10747D-0323. March 2023. https://www.masimo.com/siteassets/us/documents/pdf/plm-10747d-sales-tool-sensor-and-cable-compatibility-us.pdf. Opened 2026-11-21.
- Stryker. LIFEPAK 15 accessories guide. February 2025. Page 7. https://www.stryker.com/content/dam/stryker/ems/resources/accessory-catalogs/lifepak15_accessory_catalog.pdf. Opened 2026-11-21.
- Stryker shop. LNCS-II rainbow DCI 8-lambda SpCO adult reusable sensor, catalog number 11996-000519. https://shop.stryker.com/masimo-lncs-ii-rainbow-dci-8-spco-adult-reusable-sensor/. Related listing 11996-000520. Opened 2026-11-21.
- Zoll web store. X Series and Propaq MD/M pulse oximetry, 8000-000862 and 8000-000863. https://zollwebstore.com/defib-and-monitors/xseries-propaq-md/pulse-oximetry; 8000-000862 product page. Opened 2026-11-21.