Masimo 4002 RD SET Inf Infant Adhesive SpO2 Sensor, 20/Box

SKU: MAS-4002
The Masimo 4002 RD SET Inf is a single-patient-use adhesive wrap sensor for infants from 3 to 20 kg that measures SpO2 and pulse rate.1 It plugs into an RD SET or RD rainbow SET patient cable, not directly into the monitor. It comes in a box of 20.123 The Zoll number is 8000-001879.4 Stryker's online store lists 11171-000062 with this sensor's barcode, while Masimo's 2025 cross-reference gives that number to the CS-2 version (4472).56 Pacific Biomedical supplies the genuine Masimo 4002.

The 4002 is the RD SET counterpart of the 2328 LNCS Inf: the same 3 to 20 kg sizing, with an RD connector for RD SET cables instead of an LNCS connector for LNC cables.78

Works With:

Patient Cable 20-pin cable (LIFEPAK 15 port, see Monitors): 4072 5 ft; 4792 EMS 4 ft; 4073 12 ft
25-pin right-angle cable (LIFEPAK 35, Philips Tempus Pro): 4798 EMS 4 ft
Other RD cables: Masimo RD Patient Cables. LNC, Red LNC, M-LNC and rainbow RC cables do not take RD sensors.8910 Add the 4105 RD to LNC adapter to run RD SET sensors on an LNC-series cable.11
Monitors LIFEPAK 35: Stryker's guide lists an RD SET adhesive, Inf, as 11171-000062 (see the Stryker row)12
LIFEPAK 15: Stryker's store files 11171-000062; the 02/2025 guide lists no RD SET adhesive sensor513
Zoll R Series, Zenix: Zoll 8000-001879; the R Series takes the 14-pin MD14-05 cable, the 4080414
Stryker / Physio-Control Number Stryker's online store lists 11171-000062 with the barcode FDA registers for the Masimo 4002. Masimo's 2025 cross-reference matches that number to the CS-2 version (4472). Match the Masimo number on your box, or send it to us.563
Zoll Number 8000-001879 (RD SET Inf, Infant SpO2 adhesive sensor, 20/Box), listed for the R Series and Zenix; Zoll does not print the Masimo number, so the match rests on the name and box size.4
Replacement Tape Masimo 4046, Replacement Tapes, Inf, the RD SET tape for the Inf size; ask us for the 4046.2
Cross-Reference See the Masimo 4002 row in the Masimo Part Number Cross-Reference

Product Highlights:

  • Two infant bands: From 3 to 10 kg the preferred site is the great toe (or the toe beside it, or the thumb); from 10 to 20 kg it is the middle or ring finger of the non-dominant hand.1
  • RD connector: The sensor tab clicks into the cable with tactile and audible feedback, and only RD SET and RD rainbow SET cables take it; the LNCS Inf 2328 uses LNC cables.187
  • Care limits: Not for patients allergic to foam rubber or adhesive tape, not under added tape, and not in an MRI environment.1
  • Single patient use: Discard it after the infant, and reapply it to the same infant only while the windows are clear and the adhesive still grips.1

Specifications:

Masimo Part Number 4002
Pacific Biomedical SKU MAS-4002
Product Family RD SET Inf
Sensor Type Single-patient-use adhesive wrap sensor13
Parameters SpO2 and pulse rate1
Patient Weight 3 to 20 kg (thumb or great toe to 10 kg; finger or toe from 10 kg)1
Connector RD SET sensor connector tab, shiny contacts up18
Use Single patient use; may go back on the same patient while the windows are clear and the adhesive grips1
Pack Quantity Box of 20 sensors3
Natural Rubber Latex Not made with natural rubber latex13
Sterility Non-sterile1
GTIN 108439970095813
Directions for Use Masimo LAB-10497A1

Who Uses This:

  • Fire Departments: Companies on RD rainbow SET cables carry one adhesive size for any infant from 3 to 20 kg, with nothing to clean after the call.1
  • EMS Agencies: Ground crews wrap the great toe or thumb of a baby up to 10 kg, and a finger or toe from 10 to 20 kg, in motion and no-motion conditions.1
  • Flight and Critical Care Transport: Infant transfers between facilities use the 3 to 20 kg size on the toe, thumb or finger, with a new sensor per patient.1
  • Clinics and Urgent Care: Infant spot checks and short monitoring in the exam room, with a new sensor for each child.1

Related Products:

Frequently Asked Questions:

Which patient cable does the Masimo 4002 RD SET Inf plug into?

An RD SET or RD rainbow SET reusable cable. The 4072 and 4792 have the 20-pin monitor end, the 4798 the 25-pin right-angle end, and the 4080 the 14-pin end that Zoll sells for the R Series.8914 Push the sensor tab in, shiny contacts up, until it clicks.1

Does the Masimo 4002 fit the LIFEPAK 35, and which Stryker/Physio-Control number is it?

Stryker's LIFEPAK 35 guide lists an RD SET infant adhesive sensor, 20 per box, as 11171-000062.12 Masimo's 2025 cross-reference matches that number to the CS-2 version, the 4472, yet the barcode on Stryker's shop listing is the one FDA registers for the Masimo 4002.653

What is the Zoll number for the Masimo 4002, and does it fit a Zoll X Series?

The Zoll number is 8000-001879, Zoll's RD SET Inf, Infant SpO2 adhesive sensor, 20/Box.4 Its compatibility field reads R Series and Zenix, with no X Series entry, and Zoll prints no Masimo number, so the match rests on the name and the box of 20. For an X Series, Zoll's 3 ft Inf CS-3, the 4477, is 8000-000132.15

What weight is the Masimo 4002 for, and which digit does it go on?

It is for 3 to 20 kg. Up to 10 kg the great toe is preferred, with the toe beside it or the thumb as alternatives; from 10 to 20 kg use the middle or ring finger of the non-dominant hand, or a toe.1

Is the Masimo 4472 RD SET Inf CS-2 the same sensor as the 4002?

It has the same 3 to 20 kg range and sites, but the 4472 is the cabled version, with a 2 ft flat sensor cable that keeps the reusable patient cable away from the patient.1617

What is the difference between the 4002 RD SET Inf, the LNCS Inf 2328 and the M-LNCS Inf 2512?

Only the connector. The 4002 takes RD SET and RD rainbow SET cables, the 2328 takes LNC and Red LNC cables, and the 2512 takes M-LNC and rainbow RC cables.87

How do I put the Masimo 4002 on an infant's toe or thumb?

Run the cable along the top of the foot, set the detector on the fleshy pad of the great toe, and wrap so the emitter sits on the nail bed directly opposite it. The thumb or the toe beside it also works. Disconnect the sensor from the cable before moving it.1

Who should not use the Masimo 4002, and can it go back on the same infant?

Not for infants allergic to foam rubber or adhesive tape, and never in an MRI environment.1 It may go back on the same infant while the windows are clear and the adhesive grips; otherwise use a new sensor. Do not tape over it.

Is this a genuine Masimo 4002?

Yes. Pacific Biomedical is an authorized Masimo distributor. Every Masimo 4002 ships new in its original Masimo packaging.

How fast does the Masimo 4002 ship?

Most orders ship within 1 to 2 business days, and current availability shows on this page. Expedited and overnight shipping are available at checkout.

How do I order the Masimo 4002 or request a quote?

Add it to your cart on this page, or contact CustomerService@pacificbiomedical.com or 844.436.3526 with SKU MAS-4002. Pacific Biomedical supports net terms for agencies and hospitals and ships to all 50 states.

References
  1. Masimo. Directions for Use, RD SET Series Adt, Pdt, Inf, Neo, NeoPt and NeoPt-500 SpO2 disposable sensors. LAB-10497A. techdocs.masimo.com.
  2. Masimo. RD SET sensors brochure with ordering information (RD SET Inf 4002, Replacement Tapes, Inf 4046). PLM-10064D, 2020. masimo.co.uk.
  3. U.S. Food and Drug Administration. Global Unique Device Identification Database (GUDID), Masimo model 4002, RD SET Inf SpO2, PR, PI, PVI Infant Pulse Oximeter Adhesive Sensor, box identifier 10843997009581 and unit identifier 00843997009584. accessgudid.nlm.nih.gov.
  4. Zoll. Zoll web store, item 8000-001879, RD SET Inf, Infant SpO2 adhesive sensor, 20/Box, compatibility R Series and Zenix, reviewed October 2026.
  5. Stryker. Online store listing, Masimo RD SET Infant Adhesive Sensor, catalog 11171-000062, SENSOR,RD SET DISPOSABLE,INFANT,20-BOX, UPC 00843997009584, filed under LIFEPAK 15 and LIFEPAK 35 equipment, reviewed October 2026.
  6. Masimo. 2025 Masimo Sensor Cross Reference, rev. January 2025 (Stryker/Physio-Control and Zoll catalog numbers), with Stryker and Zoll catalog checks, reviewed October 2026.
  7. Masimo. Neonatal and infant care solutions, product information (RD SET Newborn Neonatal 4013, RD SET Inf 4002 and Neo 4003, LNCS Neo 2329). PLM-10248C, 2019. opioidhalo.masimo.com.
  8. Masimo. SET and rainbow Sensor and Cable Compatibility, with minimum software requirements. PLM-10747D, 2023. masimo.fr.
  9. Masimo. Directions for Use, RD rainbow SET Series patient cables. LAB-9000G. techdocs.masimo.com.
  10. Masimo. RD patient cables product page: RD rainbow SET MD20, R25 and RA25 cables and RD SET MD14 cables, with pin counts (masimo.co.uk). masimo.co.uk.
  11. Masimo. Directions for Use, RD to LNC adapter cable, with the RD SET sensors it takes. LAB-9182H. techdocs.masimo.com.
  12. Stryker. LIFEPAK 35 accessories guide, pulse oximetry accessories (12/2025), page 6. stryker.com.
  13. Stryker. LIFEPAK 15 accessories guide, pulse oximetry accessories (02/2025), page 7. stryker.com.
  14. Zoll. Zoll web store, item 8000-001142, RD SET MD14-05, Patient Cable, 5 FT, compatibility R Series, reviewed October 2026.
  15. Zoll. Zoll web store, item 8000-000132, Masimo RD SET SpO2 Adhesive Sensor, Infant, 3 ft, Box of 20, compatibility X Series, X Series Advanced and Zenix, reviewed October 2026.
  16. Masimo. RD SET Cabled Sensors, product information with part numbers, lengths and replacement tapes. PLM-11422C, 2022. masimo.co.uk.
  17. Masimo. RD SET single-patient-use sensors product page: RD SET cabled sensors CS-2 (2 ft), RD SET Newborn Neonatal and RD SET TFA-1 (masimo.co.uk). masimo.co.uk.
Item # Image Price Quantity

Masimo 4002 RD SET Inf Infant Adhesive SpO2 Sensor, 20/Box

MAS-4002

Masimo 4002 RD Set Disposable Infant Adhesive Sensor 20/Box. Great for precessions in 3-20kg patients.

$324.00

20-bx

9 items left

Authorized Masimo Distributor Genuine Masimo Product

Most items ship in 1 to 2 business days

Authorized Masimo Distributor
Genuine Masimo Product
Most items ship in 1 to 2 business days
Item # Image Description Quantity