Masimo 2508 M-LNCS Adtx Adult Adhesive SpO2 Sensor, 18 in, 20/Box

SKU: MAS-2508
The Masimo 2508 M-LNCS Adtx is a single-patient-use adhesive sensor for adults over 30 kg that measures SpO2 and pulse rate.1 Its 18 in lead2 ends in an M-LNCS connector that plugs straight into a rainbow RC patient cable, such as the RC-4 EMS (4481), or into an M-LNC cable, with no adapter.345 It comes as a box of 20.6 The Stryker/Physio-Control number is 11171-000039 and the Zoll number is 8000-000475.789 We supply the genuine Masimo 2508.

Works With:

Patient Cable rainbow RC cable, no adapter: 2405 RC-1, 1 ft, 2406 RC-4, 4 ft, 4481 RC-4 EMS, 4 ft.34
M-LNC cable (Masimo SET devices): 2523 M-LNC-1, 1 ft, 2524 M-LNC-4, 4 ft, 2525 M-LNC-10, 10 ft.5
Monitor Connector Philips Tempus Pro (25-pin right-angle rainbow port): the 20-pin RC cables above do not fit that port. Philips Annex B lists the rainbow RC4 RA cable (01-2088, 12NC 989803182411, Masimo 3375; ask us for the 3375), which takes M-LNCS sensors; the 25-pin right-angle 4485 RC25-4RA EMS, 4 ft is the RC25 cable we list.31011
Not For The 12 ft RC-12 (2404), which Masimo marks No for 18 in sensors, and LNC and Red LNC cables.35
Monitors LIFEPAK 15: Stryker's 2019 guide lists the adult M-LNCS adhesive sensor (11171-000039) on RC cables; the 02/2025 guide lists none, so a Red LNC bag takes the LNCS Adtx (1859).712
Zoll X Series and Propaq: Zoll's item field reads "Propaq, X Series, X Series Advanced".9
Masimo Rad-57: listed on rainbow cables.13
Philips Tempus Pro: Annex B lists only the 3 ft Adtx-3 (2509).10
Not Listed For LIFEPAK 20e and LIFEPAK 35: neither Stryker guide lists an M-LNCS sensor. Use the LNCS Adtx (1859, Stryker 11171-000019).1415
Stryker / Physio-Control Number 11171-000039. Stryker's 2019 guide lists one adult M-LNCS adhesive sensor under it, with no lead length; the cross-reference assigns it to the 18 in Adtx.78
Zoll Number 8000-000475, Zoll's M-LNCS Adtx, Adult Adhesive Sensor, 20 per box. Zoll prints no Masimo number, so the match rests on the name and the box.9
Cross-Reference See the Masimo 2508 row in the Masimo Part Number Cross-Reference

Product Highlights:

  • No adapter: The M-LNCS connector locks into a rainbow RC or M-LNC cable, and the same RC cable takes rainbow sensors when SpCO or SpMet is needed.13
  • 18 in lead: The shorter adult lead adds 1.5 ft to the cable, so a 4 ft RC-4 reaches about 5.5 ft from the monitor; the 3 ft Adtx-3 reaches about 7 ft.23
  • Site: The middle or ring finger of the non-dominant hand, or a toe; check it at least every 8 hours.1
  • Accuracy: On Masimo SET, SpO2 is within 2% Arms with no motion and at low perfusion and 3% with motion; pulse rate is within 3 bpm, or 5 bpm with motion.1

Specifications:

Masimo Part Number 2508
SKU MAS-2508
Product Family M-LNCS Adtx (Masimo SET)
Sensor Type Single-patient-use adhesive, adult16
Parameters SpO2 and pulse rate1
Patient Weight Adult, over 30 kg1
Cable Length 18 in2
Connector M-LNCS (M15), for rainbow RC and M-LNC patient cables5
Use Single patient use; reapply to the same patient while the adhesive holds1
Pack Quantity Box of 20 sensors6
Natural Rubber Latex Not made with natural rubber latex16
Sterility Non-sterile1
GTIN 10843997002506
Directions for Use Masimo LAB-6698J1

Who Uses This:

  • Fire Departments: Fire crews run it on an adult's finger through a rainbow RC cable for rehab and on-scene SpO2 and pulse rate checks.13
  • EMS Agencies: Ambulance crews run it on an adult's finger through an RC cable and reorder by Zoll 8000-000475 or Stryker 11171-000039.89
  • Flight and Critical Care Transport: Flight and critical care crews keep an adult's SpO2 and pulse rate reading through patient motion, since the sensor is labeled for motion and low-perfusion conditions in mobile environments.1
  • Clinics and Urgent Care: Clinic and urgent care staff use it for adult spot checks and short monitoring on a Masimo SET monitor with an M-LNC or rainbow RC cable.15

Related Products:

Frequently Asked Questions:

Which patient cable does the Masimo 2508 plug into, and is there an RD SET adapter?

A rainbow RC cable (2405, 2406 or 4481) or an M-LNC cable, with no adapter.345 It does not fit RD SET, RD rainbow SET, LNC or Red LNC cables, and the 12 ft RC-12 (2404) is marked not for 18 in sensors.35 RC cables need a rainbow SET device, version 7.1 or higher.3

Does the Masimo 2508 work on a LIFEPAK 15, and what are Stryker 11171-000039 and Zoll 8000-000475?

On a LIFEPAK 15 it runs on an RC cable, the route Stryker's 2019 guide lists for 11171-000039; the 02/2025 guide lists none.712 Zoll's M-LNCS Adtx, 8000-000475, matches by name and box and is listed for "Propaq, X Series, X Series Advanced".9

Masimo 2508 Adtx or 2509 Adtx-3: which should I order?

Order the one your reorder sheet matches. The 2508 has an 18 in lead and carries the Stryker and Zoll numbers in the cross-reference; the 2509 has a 3 ft lead, which Philips lists for the Tempus Pro (01-2090).2810 Both are for patients over 30 kg and ship 20 per box.1

Will the Masimo 2508 work on a LIFEPAK 35 or LIFEPAK 20e?

Not through any Stryker-listed path. The LIFEPAK 35 and LIFEPAK 20e guides list LNCS adhesive sensors, such as the adult 11171-000019, and no M-LNCS sensor.1415 The 2508 has the M15 connector, which Masimo's chart marks not for LNC or Red LNC cables.5 On those monitors use the LNCS Adtx (1859).

Can the Masimo 2508 go on a large child, or do I need the Pdtx?

The 2508 is labeled for patients over 30 kg. The pediatric Pdtx covers 10 to 50 kg, so a child from 30 to 50 kg fits either size, and a child from 10 to 30 kg takes the pediatric sensor, 2510 (18 in) or 2511 (3 ft); a child under 10 kg takes an infant sensor.1

Where does the Masimo 2508 go on a patient, and how often do I check the site?

Use the middle or ring finger of the non-dominant hand; a toe is the other labeled site.1 Place the fingertip on the dashed line with the pad over the detector, fold the sensor so the red star sits over the nail, and line up the black marks. Inspect the site at least every 8 hours, or hourly on a poorly perfused patient.1

Can I tape down or reapply a Masimo 2508 that keeps lifting?

Do not add tape: it can restrict blood flow and give inaccurate readings.1 An adult sensor can go back on the same patient while the windows are clear and the adhesive holds. Adhesive tabs come only with infant and neonatal sensors, so when it fails, use a new sensor.1

Why does the monitor show replace sensor or no reading with the Masimo 2508?

Replace the sensor when a replace sensor message shows, or when a low SIQ message stays after the monitor's low SIQ steps.1 Check that the connector is fully in and locked, the pad covers the detector, and the site is at heart level and clear of a cuff or arterial line. Nail polish, dyes and high ambient light also disturb readings.1

Is this a genuine Masimo 2508?

Yes. We are an authorized Masimo distributor. Every Masimo 2508 ships new in its original Masimo packaging.

How fast does the Masimo 2508 ship?

Most orders ship within 1 to 2 business days. Expedited and overnight shipping are available at checkout.

How do I order the Masimo 2508 or request a quote?

Add it to your cart on this page, or contact CustomerService@pacificbiomedical.com or 844.436.3526 with SKU MAS-2508. We support net terms for agencies and hospitals and ship to all 50 states.

References
  1. Masimo. Directions for Use, M-LNCS Series, LNCS Series adult, pediatric, infant, neonatal and preterm SpO2 adhesive sensors. LAB-6698J. techdocs.masimo.com.
  2. Masimo. LNCS single-patient-use sensors, product table (lead length, patient weight and replaceable wrap by model name; the M-LNCS and LNCS sensors share model names). masimo.com.
  3. Masimo. Directions for Use, rainbow Series patient cables. LAB-4523J. techdocs.masimo.com.
  4. Masimo. Directions for Use, rainbow Series patient cables, EMS. LAB-9877B. techdocs.masimo.com.
  5. Masimo. SET and rainbow Sensor and Cable Compatibility, with minimum software requirements. PLM-10747D, 2023. masimo.fr.
  6. U.S. Food and Drug Administration. Global Unique Device Identification Database (GUDID), device identifier 10843997002506, M-LNCS Adtx SpO2 Adult Pulse Oximeter Adhesive Sensor, model 2508. accessgudid.nlm.nih.gov.
  7. Stryker. Masimo sensor selection guide for SpO2 and rainbow (SpO2, SpCO, SpMet), for use with LIFEPAK 15 monitor/defibrillators. GDR 3308460_H, 2019. stryker.com.
  8. Masimo. Sensor Cross Reference (Stryker/Physio-Control and Zoll catalog numbers), rev. January 2025, with Stryker and Zoll catalog checks, reviewed October 2026.
  9. Zoll. X Series and Propaq MD/M pulse oximetry accessories, Zoll web store (M-LNCS adhesive items 8000-000475 Adtx, 8000-000476 Pdtx-3, 8000-000477 Inf-3 and 8000-000478 Neopt-3; LNCS Neo-3 8000-0323; rainbow RC-4 8000-0341 and RC-4 EMS 8000-001392). Compatibility field 'Propaq, X Series, X Series Advanced' on the M-LNCS Adtx, Pdtx-3 and Inf-3 items and on both rainbow cables. Checked October 2026.
  10. Philips. Tempus Pro user/operator manual 41-2036EN-12 (2025), Annex B.4 pulse oximetry accessories, pages 275 and 276.
  11. Philips. Masimo rainbow SET angled connector adapter cable for use with rainbow and M-LNCS SpO2 sensors, Masimo part number 3375. Product HC989803182411. usa.philips.com.
  12. Stryker. LIFEPAK 15 accessories guide, 02/2025. stryker.com.
  13. Masimo. Rad-57 Signal Extraction Pulse CO-Oximeter Operator's Manual, sensor and patient cable tables. LAB-9204A. techdocs.masimo.com.
  14. Stryker. LIFEPAK 20e genuine accessories, GDR 3300292_K. stryker.com.
  15. Stryker. LIFEPAK 35 accessories guide, 12/2025. stryker.com.
Item # Image Price Quantity

Masimo 2508 M-LNCS Adtx Adult Adhesive SpO2 Sensor, 18 in, 20/Box

MAS-2508

Disposable SpO2 finger probe for adult patients. 20/bx. PN: 2508

$218.00

20-bx

Authorized Masimo Distributor Genuine Masimo Product

Most items ship in 1 to 2 business days

Authorized Masimo Distributor
Genuine Masimo Product
Most items ship in 1 to 2 business days
Item # Image Description Quantity