Masimo 2505 M-LNCS YI Multisite Reusable SpO2 Sensor, Neonate to Adult

This item requires a clinical attestation, confirmed in your cart.

SKU: MAS-2505
The Masimo 2505 M-LNCS YI is a reusable multisite sensor for neonatal, infant, pediatric and adult patients over 1 kg that measures SpO2 and pulse rate.1 It is held on a finger, toe, foot or hand with a single-patient attachment wrap and plugs into a rainbow RC or M-LNC patient cable, not an LNC or RD cable.12 We supply the genuine Masimo 2505.

The 2505 is the M-LNCS version of the Masimo YI, also written Y-I. One set of directions covers its M-LNCS, LNCS and LNOP versions, which differ only in connector; the wrap, not the sensor, sets the patient weight. It is not for the ear: use the Masimo 2503 M-LNCS TC-I.1

Works With:

Patient Cable rainbow RC cable, no adapter: 2405 RC-1, 1 ft, 2406 RC-4, 4 ft, 4481 RC-4 EMS, 4 ft.23
M-LNC cable (SET devices): 2523 M-LNC-1, 1 ft, 2524 M-LNC-4, 4 ft.2
Not For The 12 ft RC-12 (2404), LNC and Red LNC cables, and RD cables.42
Accessories Single-patient attachment wraps in four types: CleanShield multisite, standard, standard petite and foam.1 The foam wrap is 1602, 12 per pack. Adhesive squares go on the sensor pads for every wrap except CleanShield.1
Monitors Masimo Rad-57: the manual's M-LNCS reusable sensor table includes the YI (over 1 kg), on rainbow patient cables.5
No monitor maker lists the 2505 by number; send us your monitor model.689
Stryker / Physio-Control Number None listed for the 2505. Stryker's 2019 selection guide and its LIFEPAK 15 guide name no M-LNCS YI.67
Zoll Number None listed for the 2505. Zoll's store lists no M-LNCS multisite sensor.8
Cross-Reference See the Masimo 2505 row in the Masimo Part Number Cross-Reference

Product Highlights:

  • Multisite: Two flexible sensor pads, emitter and detector, on one cable. On a finger or toe the emitter pad, whose cable is marked with a red indicator, goes on the nail side and the detector on the fleshy pad opposite it.1
  • Accuracy: On Masimo SET, SpO2 is within 2% Arms above 3 kg with no motion and in low perfusion, 3% with motion, and 3% in neonates of 1 to 3 kg; pulse rate is within 3 bpm, or 5 bpm with motion.1
  • Wrap by weight: From 1 to 10 kg the wrap is a CleanShield, standard or foam wrap, for 10 to 50 kg a petite or foam wrap, and over 30 kg any of the four; the foam wrap can be trimmed for a child's finger or toe.1
  • Not for the ear: The YI is not intended for the ear, and it is contraindicated for patients allergic to foam urethane products or adhesive tape.1
  • Site checks: Remove the sensor and inspect the site at least every 4 hours, hourly on a poorly perfused patient, and never tape the sensor in place.1

Specifications:

Masimo Part Number 2505
SKU MAS-2505
Product Family M-LNCS YI (Masimo SET)
Sensor Type Reusable multisite sensor, used with single-patient attachment wraps110
Parameters SpO2 and pulse rate1
Patient Weight Neonatal to adult, over 1 kg; wrap chosen by weight15
Connector M-LNCS (M15), for rainbow RC and M-LNC patient cables2
Use Reusable; wipe with 70% isopropyl alcohol between patients1
Natural Rubber Latex Not made with natural rubber latex110
Sterility Non-sterile110
GTIN 00843997003254
Directions for Use Masimo LAB-9123G1

Who Uses This:

  • Fire Departments: Fire crews bag the sensor with fresh wraps for the next patient and use it on a small child's toe or outer foot where a clip will not fit.1
  • EMS Agencies: EMS crews keep one YI on the ambulance for every patient size from newborn to adult, putting a fresh single-patient wrap on each patient and wiping the sensor with alcohol between patients.1
  • Flight and Critical Care Transport: Neonatal and infant transport teams put the sensor on the outer foot or palm of a 1 to 10 kg patient with a foam wrap, which fits both weight bands, for transport monitoring.1
  • Clinics and Urgent Care: Clinic and urgent care staff match the wrap to the patient's weight and use the sensor for spot checks from newborn to adult on monitors with rainbow RC or M-LNC ports.12

Related Products:

Frequently Asked Questions:

Which patient cable does the Masimo 2505 YI need?

A rainbow RC cable (2405, 2406 or 4481) or an M-LNC cable (2523 or 2524), because the 2505 has the M-LNCS connector.23 The 12 ft RC-12 (2404), LNC, Red LNC and RD cables do not take it.42 Insert the connector completely and close the cover until it locks.1

Which wrap goes with the Masimo 2505 for a neonate, child or adult?

A neonate of 1 to 3 kg and an infant of 3 to 10 kg take a CleanShield, standard or foam wrap.1 A child of 10 to 50 kg takes a petite or foam wrap, using the petite wrap's center hole from 10 to 30 kg, and a patient over 30 kg can use any of the four. The foam wrap fits every weight band.1

Can the Masimo 2505 YI go on the ear?

No. The YI is not intended for the ear. For an ear site use the M-LNCS TC-I, the 2503.1 The YI goes on a finger, toe, foot or hand, by weight.1

Where does the Masimo 2505 go on a baby or a small child?

On a neonate of 1 to 3 kg: the outer foot under the fourth toe, or the outer palm. On an infant of 3 to 10 kg: the outer foot under the fifth toe, or the palm under the fifth finger; a fat or swollen foot takes the great toe or thumb.1 Put the detector on the fleshy part and the emitter directly opposite, with a loose wrap.1

How accurate is the Masimo 2505 on neonates and adults?

In neonates of 1 to 3 kg, SpO2 is within 3% Arms with no motion, with motion and in low perfusion; above 3 kg it is within 2% with no motion and in low perfusion, and 3% with motion.1 Pulse rate is within 3 bpm, or 5 bpm with motion. The YI is indicated for both no-motion and motion conditions.1

Is the Masimo 2505 YI reusable, and what is single use?

The sensor is reusable and is wiped clean between patients; only the attachment wraps are single-patient use, so each patient gets a fresh wrap.1 Replace the sensor when the monitor shows a replace sensor message, when a low SIQ message persists across consecutive patients after troubleshooting, or when its patient monitoring time is exhausted.1

How do I clean the Masimo 2505 between patients?

Take it off the patient, remove the wrap and the adhesive squares, disconnect the cable, and wipe the sensor with a 70% isopropyl alcohol pad or mild detergent; the wrap is single-patient use.1 For low-level disinfection wipe with 1:10 bleach and water, then sterile or distilled water; never immerse the connector or sterilize the sensor.1

Is this a genuine Masimo 2505?

Yes. We are an authorized Masimo distributor. Every Masimo 2505 ships new in its original Masimo packaging.

How fast does the Masimo 2505 ship?

Most orders ship within 1 to 2 business days. Expedited and overnight shipping are available at checkout.

How do I order the Masimo 2505 or request a quote?

Add it to your cart on this page, or contact CustomerService@pacificbiomedical.com or 844.436.3526 with SKU MAS-2505. We support net terms for agencies and hospitals and ship to all 50 states.

References
  1. Masimo. Directions for Use, M-LNCS, LNCS and LNOP YI SpO2 multisite reusable sensor and single-use attachment wraps. LAB-9123G. techdocs.masimo.com.
  2. Masimo. SET and rainbow Sensor and Cable Compatibility, with minimum software requirements. PLM-10747D, 2023. masimo.fr.
  3. Masimo. Directions for Use, rainbow Series patient cables. LAB-4523J. techdocs.masimo.com.
  4. Masimo. Directions for Use, rainbow Series patient cables, EMS. LAB-9877B. techdocs.masimo.com.
  5. Masimo. Rad-57 Signal Extraction Pulse CO-Oximeter Operator's Manual, sensor and patient cable tables. LAB-9204A. techdocs.masimo.com.
  6. Stryker. Masimo sensor selection guide for SpO2 and rainbow (SpO2, SpCO, SpMet), for use with LIFEPAK 15 monitor/defibrillators. GDR 3308460_H, 2019. stryker.com.
  7. Stryker. LIFEPAK 15 accessories guide, 02/2025. stryker.com.
  8. Zoll. Zoll web store, pulse oximetry for X Series and Propaq MD/M: the M-LNCS items are the DCI (8000-000459), the E1 ear sensor (8000-000457) and adhesive sensors (8000-000475 to 8000-000478); no M-LNCS DCIP, TC-I, YI or DBI item. Reviewed October 2026.
  9. Philips. Tempus Pro user/operator manual 41-2036EN-12 (2025), Annex B.4 pulse oximetry accessories list (pages 275 and 276).
  10. U.S. Food and Drug Administration. Global Unique Device Identification Database (GUDID), device identifier 00843997003254, MasimoSET M-LNCS YI SpO2 Multisite Adult Reusable Sensor, model 2505. accessgudid.nlm.nih.gov.
Item # Image Price Quantity

Masimo 2505 M-LNCS YI Multisite Reusable SpO2 Sensor, Neonate to Adult

MAS-2505

$135.10

ea

Authorized Masimo Distributor Genuine Masimo Product

Most items ship in 1 to 2 business days

Authorized Masimo Distributor
Genuine Masimo Product
Most items ship in 1 to 2 business days
Item # Image Description Quantity