Bio-Med Devices Blender Maintenance Kit, High-Low Flow, 2008
The Bio-Med Devices 2008 is the Blender Maintenance Kit, High-Low Flow, a single preventive maintenance parts kit.1 It holds the parts used in the 2-year major overhaul of a Bio-Med Devices air/oxygen blender.2 It suits a biomedical department servicing one blender at a time.
Product Highlights:
- Single kit: Part 2008, sold as one kit; 2008B is the 10 pack3
- Overhaul parts: Filters, O-rings, check valves, diaphragms and spring2
- 2-year interval: Major overhaul, cleaning and calibration every 2 years23
- Annual step: Calibration certification every year between PMs2
- Inlet filters: Replaced during major overhauls or whenever contaminated2
What Is in the Box:
- 1 Blender Maintenance Kit1
- Filters, O-rings, check valve ball, balancing chamber balls, duck bill check valves, bleed muffler and retainer, diaphragms, spring and label. Contents are subject to change without notice.2
Specifications:
| Manufacturer | Bio-Med Devices, Inc.1 |
|---|---|
| Part number | 20081 |
| Listing name | Blender Maintenance Kit, High-Low Flow1 |
| PM interval | Every 2 years3 |
| Calibration | Annual, between PMs3 |
Size and Compatibility:
The service manual assigns part 2008 to all Bio-Med Devices blenders. The current product listing names it Blender Maintenance Kit, High-Low Flow.21
Related Products:
- Blender PM Kit, 2008B, box of 10
- High-Low Flow Blender with 3 outlet ports, 2004
- High-Low Flow Blender with 70 and 15 LPM flowmeters, 2004F7015
- Blender Hose Kit, 3 ft, 20063
- 2 in Pole Mount Bracket, 2013BH
- Oxygen Blenders collection
Product Guides and Documents:
- Repair Kits and PM Schedules page3
- Air/Oxygen Blender Service Manual, Catalog #2121 (PDF)2
- Product Listing with Part Numbers (PDF)1
Frequently Asked Questions:
What is in the 2008 blender maintenance kit?
Filters, O-rings, check valve ball, balancing chamber balls, duck bill check valves, bleed muffler and retainer, diaphragms, spring and label. Contents are subject to change without notice.2
What is the difference between 2008 and 2008B?
2008 is one kit. 2008B is the 10 pack.3
How often does an air/oxygen blender need maintenance?
A performance test before each use, calibration certification every year between PMs, and a major overhaul, cleaning and calibration every 2 years.23
What changes how often a blender needs service?
Frequency and length of use, the quality of the compressed gas sources, and environmental conditions.2
Does a blender need calibration after service?
Yes. Calibration procedures must be performed following servicing of the blender.2
When are the blender inlet filters replaced?
During major overhauls, or whenever they become contaminated. The filter element sits in the air and oxygen supply inlet fittings.2
What does an air/oxygen blender do?
It mixes medical grade air and oxygen to any concentration from 21% to 100% oxygen and delivers the mix to respiratory care devices.2
What supply pressure does a Bio-Med Devices blender use?
Air and oxygen at 50 psi, connected to two DISS fittings on the bottom of the blender. The rated supply range is 30 to 75 psi, and air and oxygen must stay within 10 psi of each other.2
Manufacturer Sources
- Bio-Med Devices. Product Listing with Part Numbers, REV060326 (accessed 2026-11-04). biomeddevices.com.
- Bio-Med Devices. Air/Oxygen Blender Service Manual, Catalog #2121, Rev 032709, archived copy of the manufacturer document (accessed 2026-11-04). frankshospitalworkshop.com.
- Bio-Med Devices. Repair Kits, Preventive Maintenance and Calibration Schedules (accessed 2026-11-04). biomeddevices.com.
Kit contents are from Bio-Med Devices Service Manual Rev 032709 and are subject to change without notice. Factory installed gas supply fittings must not be substituted with parts not approved by Bio-Med Devices.2