Bio-Med Devices Neo2 Low Flow Blender, 15 and 3.5 LPM, 2003FF

SKU: BMD-2003FF

The Bio-Med Devices 2003FF is a Low Flow Neo2 air/oxygen blender with two flowmeters, 0 to 15 lpm on the left and 0 to 3.5 lpm on the right.1 The Neo2 Blend series is the Bio-Med Devices model for the NICU, and the left flowmeter is placed for resuscitation bag connection.2

Product Highlights:

  • Two flowmeters: 0 to 15 lpm left and 0 to 3.5 lpm right, green O2 labels1
  • Integrated bleed: Rotating the right flowmeter activates and deactivates the bleed1
  • NICU model: Neo2 Blend series, 7 configurations2
  • FiO2 accuracy: 0 to 3 lpm with bleed and 3 to 30 lpm without3
  • Therapies: Oxygen administration, bubble CPAP and HFNC1

Specifications:

Manufacturer Bio-Med Devices, Inc.4
Part number 2003FF1
Series Low Flow, Neo2 Blend4
Flow range 0 to 30 lpm1
Left port 0 to 15 lpm flowmeter, green O2 label1
Right port 0 to 3.5 lpm flowmeter, green O2 label1
Bleed 3 lpm for flows under 3 lpm1
Fittings DISS input and output3

Size and Compatibility:

The 2003FF1 swaps the right flowmeter for 1 lpm, and the 2003FF15 runs two 15 lpm flowmeters.5

Related Products:

Product Guides and Documents:

Frequently Asked Questions:

What flowmeters does the 2003FF have?

0 to 15 lpm on the left and 0 to 3.5 lpm on the right, both with green O2 labels.1

What is the difference between 2003FF, 2003FF1 and 2003FF15?

The flowmeters. 2003FF has 3.5 and 15 lpm, 2003FF1 has 1 and 15 lpm, and 2003FF15 has two 15 lpm.5

What is an oxygen blender used for?

Oxygen administration, bubble CPAP therapy and HFNC therapy.1

What is bleed flow on an air/oxygen blender?

A blender needs a minimum flow of gas passing through it to stay accurate. The bleed supplies that flow by releasing gas to atmosphere.6

How is the blender bleed turned on and off?

The right side of the blender controls it. A DISS fitting turns the bleed on when something is connected. A knob or flowmeter must be pushed in to turn; a flowmeter turns the bleed on when rotated to vertical.6

What is the normal range for an oxygen flow meter?

On the 2003FF the flowmeters read 0 to 15 lpm and 0 to 3.5 lpm.1

Is the 2003FF MRI safe?

No. The FDA device record lists 2003FF as MR Unsafe.7 Bio-Med Devices lists separate MRI blenders, including 2003M.4

What fittings does the blender use?

DISS input and output port fittings. For NIST input fittings, add part 2007 to the order.3

How often does an air/oxygen blender need maintenance?

A performance test before each use, calibration certification every year between PMs, and a major overhaul, cleaning and calibration every 2 years.89

Manufacturer Sources
  1. Bio-Med Devices. Blender Gallery Product Sheet, DDOC063 Rev031925 (accessed 2026-11-04). biomeddevices.com.
  2. Bio-Med Devices. Air/Oxygen and Neo2 Blenders page (accessed 2026-11-04). biomeddevices.com.
  3. Bio-Med Devices. 2003FF Blender Spec Sheet, DDOC071 Rev120925 (accessed 2026-11-04). biomeddevices.com.
  4. Bio-Med Devices. Product Listing with Part Numbers, REV060326 (accessed 2026-11-04). biomeddevices.com.
  5. Bio-Med Devices. 2003FF blender page (accessed 2026-11-04). biomeddevices.com.
  6. Bio-Med Devices. The Mystery of Blender Bleeds Unraveled (accessed 2026-11-04). biomeddevices.com.
  7. FDA AccessGUDID. Device record 2003FF, DI 00813841020201 (accessed 2026-11-04). accessgudid.nlm.nih.gov.
  8. Bio-Med Devices. Air/Oxygen Blender Service Manual, Catalog #2121, Rev 032709, archived copy of the manufacturer document (accessed 2026-11-04). frankshospitalworkshop.com.
  9. Bio-Med Devices. Repair Kits, Preventive Maintenance and Calibration Schedules (accessed 2026-11-04). biomeddevices.com.

Specifications are from Bio-Med Devices documents. Use only as directed in Bio-Med Devices labeling.4