AirLife air-Q Disposable Removal Stylets, named Air-Q3 Exchangers by the manufacturer, are accessories designed exclusively for the Air-Q3, supporting intubation procedures through a supraglottic airway.
2 After an oral endotracheal tube is placed through the Air-Q3, the stylet holds the tube in place while the Air-Q3 is slowly withdrawn over it.
3 Pick the stylet that matches the Air-Q3 size in use; each size has its own page below.
Product Highlights:
-
Three sizes: Stylets #00, #0 and #1 match Air-Q3 sizes 0 to 1.5, 2 to 3 and 4 to 5.1
-
Max tube: Maximum oral ETT 4.5, 6.0 and 9.0 mm by stylet size.1
-
Purpose: Designed exclusively for the Air-Q3 to support intubation through a supraglottic airway.2
-
Latex: AccessGUDID lists no natural rubber latex.4
-
MRI: AccessGUDID lists the stylets as MR Safe.4
What Is in the Box:
- The manufacturer table lists each stylet size at PK 10.1
- AccessGUDID lists a package of 15 for each stylet size.4
- Each size page on this store is for a single stylet.
Specifications:
| Manufacturer |
AirLife4
|
| Brand |
Air-Q (Air-Q3 Exchanger)2
|
| Part number (MPN) |
10-1006 and 10-1005 (10-1004 not sold here)1
|
| Size |
Stylet #00, #0 and #11
|
| Pack |
PK 10 per manufacturer table1
|
| Sterility |
Non-sterile4
|
| Use |
Single use4
|
| MRI |
MR Safe4
|
| Latex |
Not made with natural rubber latex4
|
| Storage |
-20 C to 50 C; keep dry and away from sunlight4
|
Size and Compatibility:
| MPN |
Stylet size |
Air-Q3 size |
Max oral ETT |
Patient weight |
PK |
| 10-1006 |
00 |
0 to 1.5 |
4.5 mm |
0 to 17 kg |
101
|
| 10-1005 |
0 |
2 to 3 |
6.0 mm |
17 to 60 kg |
101
|
| 10-1004 |
1 |
4 to 5 |
9.0 mm |
over 60 kg |
101
|
Size #1 (10-1004) is listed for reference and is not sold from this page.1
Related Products:
Product Guides and Documents:
Frequently Asked Questions:
Which removal stylet do I need?
Match the Air-Q3 size: #00 (10-1006) for sizes 0 to 1.5, #0 (10-1005) for sizes 2 to 3 and #1 (10-1004) for sizes 4 to 5.1
What is an Air-Q airway?
The Air-Q3 is a supraglottic airway engineered for quick, effective airway placement in routine and challenging cases; its wide airway conduit and flexible design support intubation and ventilation.2
What is an Air-Q3 Exchanger?
It is the manufacturer name for the removal stylet: an accessory designed exclusively for the Air-Q3 that supports intubation through a supraglottic airway.2
What is the largest tube each size accepts?
Maximum oral ETT 4.5 mm for #00, 6.0 mm for #0 and 9.0 mm for #1.1
How is the Air-Q3 removed over the stylet?
Per the user guide: remove the OETT connector and insert the tapered end of the removal stylet into the proximal OETT until the adapter fits. Deflate the cuff and pilot balloon, then, with an inward stabilizing force on the stylet, slowly withdraw the Air-Q3 outward over the rod. Reposition the OETT, replace its connector and reconnect the breathing device.3
Are the removal stylets sterile?
No. AccessGUDID lists them as non-sterile and single use.4
Are they MRI safe?
AccessGUDID lists the stylets as MR Safe.4
How should they be stored?
AccessGUDID lists storage from -20 C to 50 C, kept dry and away from sunlight.4
Manufacturer Sources
- SunMed (AirLife). Air-Q3 Supraglottic Airways literature sheet, LIT-20014-C1 Rev 3, Disposable Removal Stylets table. connect.myairlife.com.
- AirLife. Air-Q3 product page, Air-Q3 Exchangers section (accessed 2026-11-01). myairlife.com.
- SunMed (AirLife). Air-Q3 User Guide, UG-20000-C1 Rev 1, removal of the Air-Q3 over the removal stylet. connect.myairlife.com.
- U.S. FDA AccessGUDID. AirLife Air-Q stylet size #0, catalog 10-1005, device identifier 10889483214114; the size #00 record (10-1006, 10889483214121) lists the same fields (accessed 2026-11-01). accessgudid.nlm.nih.gov.
Rx only. FDA AccessGUDID lists the Air-Q stylets for prescription use and single use. Read the AirLife labeling and user guide before use.